How do I reset the Baxter 6201 to factor settings? Trying to get the slide clamp alarm off!!! Thanks
You press and hold the lock panel on the back of the pump then you press and hold the stop botton, then you turn on the pump.
After you get on to Configuration mode, you switch to different settings using the "Next" button. Anytime you change or enter a new value you have to press the green "start" button to save it.
Good luck
Jason
Once I have the clamp alarm up, how do you scroll to "off" position?