EUR
en
The proper operation of a diaphragm pump often depends on its proper installation and careful maintenance. However, even a properly installed pump can sometimes have problems, such as priming, pressure, pulsation, noise or oil consumption problems. If you feel that the pump is not performing as it should, you must take immediate action to understand the cause of malfunctioning and restore it. In this article we have summarized the most common problems that may occur during the normal use of a diaphragm pump and the necessary actions to fix them. Some procedures are routine maintenance, while others should be carried out by skilled technicians or at authorized service centers.
When the pump does not prime or draw water, there are generally three causes: The control regulator is in "pressure", Problems on the suction line, i.e. pipes or fittings are sucking in air, Clogged suction filter.
Pressure problems may be caused by an RPM speed that is not adequate to the technical characteristics of the pump, in this case it is sufficient to reset the number of revolutions indicated in the manual to recreate the pressure necessary for a correct use. If the pressure continues to remain low or equal to zero, the problem may probably concern the nozzles; if the nozzles are worn or with a flow rate exceeding that which can be reached by the pump, they must be replaced.
Irregular pressure can also have several causes: First check the suction line, if pipes or fittings are sucking air the pressure inside the pump is obviously affected. The pulsation dampener may be set incorrectly. Another cause may be the incorrect configuration of the pressure regulator, check the pressure setting of the regulator (and if necessary, repair or replace it).
In the event of excessive pulsations or vibrations, you should certainly check the pulsation dampener, as the purpose of this instrument is to reduce pulsations. Every 50 hours of work you should check the air load inside and if necessary reset pressure by means of a pump.
If the pump starts becoming noisy, check the oil level immediately, if it has decreased it means that the suction line is clogged. Just inspect the suction line (especially the filter) and remove any obstruction.
If excessive oil consumption is associated with a milky color (of same) it is very likely that one or more diaphragms are broken or damaged. The diaphragms separate the pumping chamber from the transmission, preventing the pumped fluid from coming into contact with the mechanical parts and the oil; when the diaphragm breaks, the fluid filters in the oil making it milky. At this point it is important to replace all diaphragms and oil to avoid major damage.
There are several causes to take into account for this type of problem: As usual, the first thing to do is to check if the suction line is clogged and clean the filter. One of the most common causes is cavitation, vapor bubbles form inside the fluid which implode and ruin the diaphragm. If the liquid is too thick, use a water-based liquid. Excessive speed may also damage the diaphragms, it is good practice to keep below 550 RPM. Always check that the suction valve is closed correctly and that the liner holes are positioned correctly. In addition, the diaphragm may also be affected by chemical aggression, always check the compatibility between the diaphragm and the chemical product used.
This is a breakage due to wear, when the diaphragm is used beyond the permitted time (300 hours) the retention disc causes a friction and then lacerations. It is good practice to change the diaphragms every 300 hours or at the beginning of every season (the shorter of the two).
This cut is produced by the presence of air in the oil bath under the diaphragm. The air was probably not purged when assembling or repairing, and the diaphragm must now be replaced, taking care to purge the air from the pump.
This damage can have several causes: Lack of oil between piston and liner, to be filled up, Excessive suction pressure, Pump speed too low, Incorrect position of the liner holes, Delivery valve open.
In this case the deformation is caused by chemical incompatibility between diaphragm and chemical product.
Bookmark
Daniel Féau processes personal data in order to optimise communication with our sales leads, our future clients and our established clients.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.